Sign inSubmit
ArticleArticle metrics
Research Article · Open access · Published 16 December 2016

Biofidel FEA Modeling of Customized Hybrid Biological Hip Joint Prostheses, Part I: Biomechanical Behavior of Implanted Femur

Raffaella Aversa1, Florian Ion T. Petrescu2, Relly Victoria V. Petrescu2 & Antonio Apicella1

American Journal of Biochemistry and Biotechnology, Volume 12, Issue 4, pp. 270–276 · doi.org/10.3844/ajbbsp.2016.270.276

8,716 accesses4,416 downloads2 citations (Crossref)Journal metrics
Cite Check for updates

Abstract

Biofidel femur Finite Element Models have been developed using specific combination of Computer Tomography segmentation and solid modeling software tools able to represent bone physiology and structural behavior. These biofidel Finite Element Models (FEM) is used to evaluate the modification of the physiological stress distribution in a prosthesized femur and to assess new design criteria for the development of biomimetic hybrid biological hip prostheses. The faithful models proposed allowed us to properly consider the not isotropic characteristics of the proximal epiphysis of the femur and for the isotropic behavior in diaphysis to explain the critical alterations of the stress distribution in a resected femur following the implantation of a traditional hip joint prostheses. It has been shown that a wide region of the femur diaphysis is completely shielded by the rigid prosthesis significantly altering the physiological stress distribution that should guaranty a healthy bone growth and regeneration.

Licence and reuse

Open access. This article is licensed under a Creative Commons Attribution 3.0 Licence (CC BY 3.0), as printed on the article, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original authors and the source. No permission is needed.

How to reuse this article

Responses (0)

Have a comment on this research? Sign in to respond. Responses are moderated and published with your name and affiliation; the authors are invited to reply. How we use your dataWrite a response

About this article

Cite this article

Aversa, R., Petrescu, F. I. T., Petrescu, R. V. V. & Apicella, A. (2016). Biofidel FEA Modeling of Customized Hybrid Biological Hip Joint Prostheses, Part I: Biomechanical Behavior of Implanted Femur. American Journal of Biochemistry and Biotechnology, 12(4), 270–276. https://doi.org/10.3844/ajbbsp.2016.270.276

Licence
CC BY 3.0
Issue
Volume 12, Issue 4
Policies
Editorial independenceUse of generative AIData availabilityAppeals and complaints
Version
Version of record
Biofidel FEA Modeling of Customized Hybrid Biological Hip Joint Prostheses, Part I: Biomechanical Behavior of Implanted FemurAmerican Journal of Biochemistry and Biotechnology · Volume 12, Issue 4 · pp. 270–276 · CC BY 3.0Download