A Comparative Study of the Ovine Haemogram: Cell-Dyn 3500 Versus Manual Methods

Labrini V. Athanasiou1, Constantinos G. Giannakopoulos2, Zoe S. Polizopoulou3, Eleni Katsogiannou1 & Vassilis I. Kontos4

American Journal of Animal and Veterinary Sciences, Volume 8, Issue 4, pp. 203–209 · doi.org/10.3844/ajavsp.2013.203.209

7,179 accesses6,972 downloads2 citations (Crossref)Journal metrics
Cite Check for updates

Abstract

Manual hematology techniques traditionally applied in farm animal medicine are time consuming and labor-intensive, especially when large numbers of samples have to be processed. As a result, several automated hematology instruments have been developed for use in these species. An automated hematology analyzer (Abbott Cell-Dyne 3500 system) was used in performing a complete blood cell count and differential counts of white blood cells in sheep blood samples. The system was compared with basic manual hematologic techniques. A linear regression was used to assess correlation between the two methods. Correlation coefficients (R2) were good for the hematocrit, the total white blood cell count, the neutrophils, the lymphocytes and the platelet count, while a poor correlation existed in monocytes, eosinophils and basophils. The automated and the manual technique were also compared in terms of sensitivity using the Sensitivity Ratio (SR). The automated analyzer was slightly more sensitive than the manual technique for all parameters tested except for monocytes, eosinophils and basophils were the difference was greater and the automated analyzer was 1.5, 2.5 and 2 times more reliable, respectively. The method bias was also calculated. It seems that the overall performance of the automated analyzer justifies its utilization in sheep blood analysis, although as for any analyzer used in any species, a stained blood film evaluation remains an indispensable technique to confirm the results being reported by the automated analyzer and provide additional information for the ovine haemogram.

Licence and reuse

Open access. This article is licensed under a Creative Commons Attribution 3.0 Licence (CC BY 3.0), as printed on the article, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original authors and the source. No permission is needed.

How to reuse this article

Responses (0)

Have a comment on this research? Sign in to respond. Responses are moderated and published with your name and affiliation; the authors are invited to reply. How we use your dataWrite a response

About this article

Cite this article

Athanasiou, L. V., Giannakopoulos, C. G., Polizopoulou, Z. S., Katsogiannou, E. & Kontos, V. I. (2013). A Comparative Study of the Ovine Haemogram: Cell-Dyn 3500 Versus Manual Methods. American Journal of Animal and Veterinary Sciences, 8(4), 203–209. https://doi.org/10.3844/ajavsp.2013.203.209

Licence
CC BY 3.0
Issue
Volume 8, Issue 4
Policies
Editorial independenceUse of generative AIData availabilityAppeals and complaints
Version
Version of record
A Comparative Study of the Ovine Haemogram: Cell-Dyn 3500 Versus Manual MethodsAmerican Journal of Animal and Veterinary Sciences · Volume 8, Issue 4 · pp. 203–209 · CC BY 3.0Download